Tuttnauer EZ10P Fully Automatic Autoclaves with Printer
In Stock & Ready to Ship
All orders are shipped within 3-5 business days. Product ETA 7-10 business days.
This item is only sold to professionals
Introducing the EZ10P, the epitome of efficiency and convenience in steam sterilization. This compact, fully automatic autoclave is engineered to simplify your sterilization process with a single button that initiates a complete cycle—from steam penetration to post-vacuum drying—until your equipment is sterilized, dried, and ready for use.
Efficient and User-Friendly Operation: The EZ10P boasts a bright, easy-to-read control panel that ensures thorough steam penetration and efficient post-vacuum drying, all within a closed-door system to maintain sterility. Equipped with replaceable 0.2μm Hepa air filters, it guarantees clean air circulation throughout the chamber, enhancing the sterilization quality. Customizable program parameters can be easily adjusted and stored according to your specific needs.
Key Features:
- One-Button Activation: Simplifies the sterilization process, making it user-friendly and time-efficient.
- Closed Door Drying System: Ensures packs and pouches are dried efficiently while maintaining sterility.
- High Efficiency Air Pump: Circulates air in the chamber during the drying cycle for superior drying results.
- Standard Internal Printing: Automatically documents the date, time, temperature, and pressure for each cycle.
- Easy-to-Read Control Panel: Features program selection, a detailed digital display, programmable keys, and system indicators for effortless operation.
- Customizable and Versatile: Allows for the customization of program parameters and includes pre-loaded cycles for unwrapped instruments, wrapped/packs, glassware, and an additional drying cycle to meet various sterilization needs.
Whether it's for unwrapped instruments, wrapped packs, or glassware, the EZ10P autoclave delivers outstanding sterilization and drying performance, making it an indispensable tool for any healthcare or laboratory setting seeking efficiency, reliability, and ease of use.
Sterilization Program
A tailored program for each requirement
Comes equipped with preset cycles for unwrapped instruments, wrapped/packs, glassware, plus an extra drying cycle.
Standard unwrapped cycle times:
- Hot start at 14 minutes
- Cold start at 30 minutes
Data & Software
Built-in printer for cycle documentation
- Date and time
- Temperature
- Pressure
- System messages and alerts
Technical Specs
Overall Dimensions (DxWxH) | 21.5” x 20” x 14.4” 545mm x 510mm x 365mm |
Chamber Volume | 23 Liter 6 Gallon |
Chamber Dimensions | 10” x 19” 254mm x 475mm |
Tray Capacity | 4 Trays |
Tray Dimensions | 16.3” x 6.7" x 0.6” 415mm x 170mm x 20mm |
Standard Cassette Capacity | 3 full & 3 half* |
Voltage & Frequency | 120V 50-60Hz |
Power & Current | 1400W 12A |
Shipping Weight | 95 Lbs. 43 Kg. |
Warranty | 2 years parts and labor |
ISO 13485:2003
ISO 9001:2000
Installation
Installation of Tabletop Autoclaves
Just the essential installation steps.
For units with a Microprocessor version # ending in WP (the version # is displayed on the screen each time the unit is turned on).
- Place unit on a sturdy level counter.
- Fill the Reservoir with distilled water
- The unit is ready to operate
To Adjust the Filling Time:
- Press the STOP Key repeatedly until the message “Code:xxx” appears.
- Using the UP/DN arrow keys change the code to 105, then press the STOP Key.
- A message will be displayed saying “Water in = xx sec”
- Using the UP/DN arrow keys change the seconds according to the following:
- EZ9 = 30 sec
- EZ10 = 35 sec
- 3870EK = 65 sec
- On some older units the message “Ea Type:” will appear,using the UP/DN arrow keys select either “0” for an E or EK type unit or “1” for an EA or EKA type unit.
- Press the STOP Key to finish
- The unit is ready to operate
For units with a Microprocessor version # NOT ending in WP
- Follow steps 1–5 for Manual Units above,then empty the water from the chamber and check the automatic water fill.
- Turn the unit on
- With the door open press and hold the door switch,then press the start button. When water starts flowing release the door switch.
- Water should come up to the same spot as the measured amount had.
- If need be adjust the automatic fill using the following procedure.
To Adjust the Automatic Filling:
- While leaving the door open turn power off.
- Press and hold the Manual Water Fill Button (this is the button with the two arrows).
- Turn the power back on
- When the normal display screen appears release the fill button –immediately press it in again and monitor the water flow
- Continue to hold the button until water reaches groove at front –then release button – wait ten seconds – the unit is reprogrammed.